TYPE | OLT-1/2/3/7/10BG | |
Total Volume | 1/2/3/7/10 | |
Vessel | 316L Stainless Steel +Borosilicate Glass | |
Stirrer | Magnetically Driven(Cell lift impeller Marine Blade impeller pitched blade impeller for option) | |
Sterilization | In Autoclave | |
Basic Features | Temperature,RPM,PH,DO,Antifoam,Feed,Aeration,Pressure,4-Gas system | |
Advanced Features | Level,Weight Feed,Multi-Feeds,Mehanol&Ethanol,Exhaust O2&CO2 | |
Controller | OLT-7000 Technical parameter |
| |
Parameter | Guideline | Controller data processing function | |
pH | Displaying range: 0.00~14.00±0.01 | pH value historical and trend curve analysis Acid, alkali dosage curve analysis | |
Antifoam | Automatic PID control and alarm | Antifoam value historical and trend curve analysis Foam error state record | |
Rpm | Automatic enactment and control(50~1000rpm±5‰) | Rpm value historical and trend curve analysis Batch report form analysis Keep and deliver data | |
Temperature | Temperature of cooling water: +5℃-65℃±0.2℃ | Temperature value historical and trend curve analysis Batch report form analysis Keep | |
Feeding | PID automatic enactment and control | Feeding value historical and trend curve analysis | |
Pressure | Manual control, instrument display |
| |
Air flux | Manual control, instrument display | ||
Performance parameter | ||
Equipment | Item | Reqirements and manufacturers |
. | reactor body | glass and strainless steel cylinder, 5L litres total volume, 2.5~4 litres working volume, sterilizable in situ, magnetic force stirrer |
Aeration system | include pressure control valve, rotameter, aeration filter(0.2μm) | |
Agitation system | bottom magnetic force stirrer, AC motor, ca.50~1200rpm,±1% | |
Temperature control | double walled dished base for temperature control, cooling water temp.+5~65℃,±0.5℃ | |
Anti foam control | antifoam probe check and add antifoam agent by one peristaltic pump | |
Feeding control | feeding by one peristaltic pump,≥3.7ml/min | |
pH control | PID control of two pumps to add acid or alkali, pH display range:0~14,±0.01, pH electrode: Swiss Mettler-Toledo | |
DO control | DO display range:0~150%,±1, DO electrode: Swiss Mettler-Toledo | |
Control system | The essential part of OLT-7000 consists of10.5"touch prainter and PLC. The standard display is used as showing interface and the keyboard and mouse can be used to operate. One controller can control 2~12 fermenters at the same time and | |



Our Advantages